5 Operation
5.1 Acquiring the X-ray image
H
50
14. Press the forehead support adjustment key (H) and the temple
support adjustment key (I).
Ä The forehead and temple supports stop moving automatically
when they come into contact with the patient's head.
Ensure that the patient's head does not move backward when
the forehead support is put in place.
15. Check the patient's position and make any final corrections as
necessary.
16. Swivel the control mirror back into place. Press the right recess on
the touchbar D.
Ä The patient can see himself in the control mirror.
17. Instruct the patient to exhale, place his tongue against the roof of
his mouth, and hold this position until the end of the exposure.
Ä The patient is positioned in the unit.
I
IMPORTANT
The slice width is selected automatically for different dental arches
with the temple support setting, and the radiation time is also changed
through this in accordance with the temple support width which is set.