With the two-part exposure program TM 1 two exposures (TM 1.1 and
TM 1.2) are prepared.
ü The first temporomandibular joint exposure has been released as
kV
mA
described above. The exposure program has changed from TM 1.1
to TM 1.2 on the digital display of the Multipad. The rotating unit has
automatically moved to the starting position.
1. Instruct the patient to open his/her mouth.
Ä The patient has opened his/her mouth without changing his/her
position.
2. Press the release button A and hold it down until the end of the
second exposure.
Ä The second exposure is released.
3. Wait until a short pulsed tone sequence sounds (can be deactivated
by a service engineer).
Ä The exposure is finished when a row of dots and the program
number alternately appear on the digital display of the Multipad.
The end of the exposure cycle can also be seen on the
SIDEXIS monitor, namely when the progress indicator shows
100% and the preview image starts to build up. The forehead
support and temporomandibular joint supports open
automatically.
4. Let go of release button (A).
Ä The second exposure is completed. Proceed as described
above starting with step 3.
5 Operation
5.1 Acquiring the X-ray image
83