Dentsply Sirona inFire HTC speed Operating Instructions Manual

Dentsply Sirona inFire HTC speed Operating Instructions Manual

Hide thumbs Also See for inFire HTC speed:
Table of Contents

Advertisement

09.2017
New as of:
inFire HTC speed
Operating Instructions
English
Operating instructions
=

Advertisement

Table of Contents
loading

Summary of Contents for Dentsply Sirona inFire HTC speed

  • Page 1 09.2017 New as of: inFire HTC speed Operating Instructions English Operating instructions...
  • Page 2: Table Of Contents

    Contents Dentsply Sirona Operating Instructions inFire HTC speed Contents Dear customer,......................General data......................Structure of the document................2.1.1 Identification of the danger levels............. 2.1.2 Formats and symbols used .............. Warranty......................Safety information ....................Technical description....................Description of the furnace ................
  • Page 3 Non-precious metal (NPM) sintering (optional) ..........6.5.1 Safety instructions ................6.5.1.1 Basic information .............. 6.5.1.2 Use of argon in the inFire HTC speed with gas management module 6.5.2 Scope of supply................6.5.2.1 Retrofitting a gas management module......6.5.3 Enhanced program list ..............
  • Page 4 Contents Dentsply Sirona Operating Instructions inFire HTC speed Maintenance......................10.1 Changing fuses F1/F2 ................... 10.2 Replacing the battery (only for devices with serial numbers < 5000)>..10.3 Furnace chamber insulation ................Malfunctions ......................11.1 Error messages from electronic components..........11.2 Miscellaneous errors and their causes............
  • Page 5: Dear Customer

    1 Dear customer, Operating Instructions inFire HTC speed Dear customer, Thank you for purchasing your inFire HTC speed high-temperature furnace from Sirona. It will support you with your work for many years, as it was developed and built with state-of-the-art technology.
  • Page 6: General Data

    2 General data Dentsply Sirona 2.1 Structure of the document Operating Instructions inFire HTC speed General data Please read this document completely and follow the instructions exactly. You should always keep it within reach. Original language of the present document: German. Structure of the document 2.1.1...
  • Page 7: Formats And Symbols Used

    Dentsply Sirona 2 General data Operating Instructions inFire HTC speed 2.2 Warranty 2.1.2 Formats and symbols used The formats and symbols used in this document have the following meaning: Prerequisite Prompts you to do something. 1. First action step 2. Second action step Alternative action ➢...
  • Page 8: Safety Information

    CAUTION Risk due to improper operation The inFire HTC speed high-temperature furnace may only be operated by employees who are familiar with the content of these operating instructions. Signs and labels on the laboratory furnace must be kept in legible condition at all times.
  • Page 9: Dentsply Sirona

    CAUTION Risk due to incorrect setup The inFire HTC speed high-temperature furnace may only be set up in dry rooms and must never be in contact with any liquids. Furniture items and other equipment in the vicinity of the furnace must not contain any explosive, flammable or easily ignitable materials.
  • Page 10: Dentsply Sirona

    Operating Instructions inFire HTC speed WARNING Risk of burns due to hot surfaces The inFire HTC speed high-temperature furnace may develop hot surfaces in some areas after extended use. Use the necessary caution around the unit, especially when the furnace is operated for extended periods.
  • Page 11: Technical Description

    Dentsply Sirona 4 Technical description Operating Instructions inFire HTC speed 4.1 Description of the furnace Technical description Description of the furnace The heating system consists of six high-quality MoSi heating elements. Thanks to the excellent interior insulation, the energy consumption of the unit is low.
  • Page 12: Certification

    Further products which are connected to this unit must also bear the CE mark. These products must be tested according to the applicable standards. We declare conformity for the inFire HTC speed ceramic sintering furnace on the basis of the following standards: ● Safety: IEC 61010-1:2001 2nd edition and IEC 61010-2-010:2003 2nd edition ●...
  • Page 13: Technical Data

    Dentsply Sirona 4 Technical description Operating Instructions inFire HTC speed 4.4 Technical data Technical data Model designation: inFire HTC speed Type of protection against electric shock: Protection Class I device Degree of protection against ingress of water: Ordinary device (without protection against ingress of...
  • Page 14: Setup

    5 Setup Dentsply Sirona 5.1 Installation site Operating Instructions inFire HTC speed Setup Installation site The unit is designed for desktop use and requires a level footprint of approx. 500 x 600 mm with a loading capacity of 80 kg. The unit must not be installed at sites with a high level of humidity or dust! ➢...
  • Page 15: Electrical Connection

    Dentsply Sirona 5 Setup Operating Instructions inFire HTC speed 5.2 Electrical connection Electrical connection Building installation The following electrical installation requirements are to be fulfilled for the high-temperature inFire HTC speed furnace: ● The furnace requires a separate electric circuit. ● The electrical building installation must be protected by a type C circuit breaker with a rating of at least 16 A.
  • Page 16: Dentsply Sirona

    5 Setup Dentsply Sirona 5.2 Electrical connection Operating Instructions inFire HTC speed RS232 interface A type RS232 serial interface (A) is located on the rear side of the unit. The use of this interface is restricted to the service engineer and is described in the service manual.
  • Page 17: Operation

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.1 Controls and displays Operation Controls and displays The controller is equipped with state-of-the-art microprocessor technology, which allows for processing a great variety of heating curves with utmost precision. Operation of the unit is menu-driven via a membrane keyboard and an LCD display.
  • Page 18: Key Functions

    6 Operation Dentsply Sirona 6.1 Controls and displays Operating Instructions inFire HTC speed 6.1.1 Key functions Symbol Designation Function Start/stop button starts/stops the selected program Load program key loads the program from memory Memory key saves a program to memory Casting time key For setting the casting time;...
  • Page 19: Switching On The Unit

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.2 Switching on the unit Description of the columns on the left side with the freely programmed furnace programs Column Designation Description Heating rate (heating and The heating and cooling down speed of the corresponding level is cooling down speed) set up / displayed in this column.
  • Page 20: Initial Startup

    6 Operation Dentsply Sirona 6.3 Practical use Operating Instructions inFire HTC speed 6.3.1 Initial startup Burning-in sintering aids Tip: We recommend subjecting all sintering aids (sintering tray, sintering shuttle) to a gradual burning-in cycle before sintering for the first time. To do this, load the furnace with the sintering aids and start the "inCoris ZI classic"...
  • Page 21: Preparing The Furnace Door

    6.3 Practical use 6.3.2.2 Preparing the furnace door 6.3.2.2.1 Furnace door with serial numbers < 5000 Furnace door components of the inFire HTC speed with serial numbers < 5000 Door insulation Door insulation with sintering pearls 61 30 459 D 3497...
  • Page 22: Dentsply Sirona

    6 Operation Dentsply Sirona 6.3 Practical use Operating Instructions inFire HTC speed Furnace door components of the inFire HTC speed with metal sintering option for serial numbers < 5000 Replacement door for metal sintering option Design with sintering bells 61 30 459 D 3497...
  • Page 23: Dentsply Sirona

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.3 Practical use 6.3.2.2.2 Furnace door with serial numbers ≥ 5000 Overview of individual components ● Upper door insulation brick ● Split support disc ● Speed sintering tray ● Stacking disc ● Small centering tube ●...
  • Page 24: Dentsply Sirona

    6 Operation Dentsply Sirona 6.3 Practical use Operating Instructions inFire HTC speed Constructing the furnace door 1. Fit the lower door insulation brick A using the small ceramic tube B and place it in the door panel. 2. Position the upper door insulation brick on the small ceramic tube on the lower door insulation brick.
  • Page 25: Dentsply Sirona

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.3 Practical use Using a second sintering tray If you wish to stack, you require a stacking disc that is not split. ✔ The first sintering tray is constructed as described above. NOTICE...
  • Page 26: Program Selection

    6 Operation Dentsply Sirona 6.4 Operating modes Operating Instructions inFire HTC speed 6.3.3 Program selection 1. Press the Load Program key.  The LOAD PROGRAM screen is displayed. The program name is displayed on the left side and the corresponding program number on the right side.
  • Page 27: Performing Super Speed Sintering

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.4 Operating modes 6.4.1.1 Performing super speed sintering Auxiliary tools & equipment ● Super speed crucible fork ● Super speed crucible ● Superspeed fireproof tray ● Superspeed covering Starting the super speed program 1.
  • Page 28: Dentsply Sirona

    6 Operation Dentsply Sirona 6.4 Operating modes Operating Instructions inFire HTC speed 1. Start the super speed process by pressing the lift button.  The loading sequence lasts approx. 15 seconds from the end position of the furnace door. Loading must be completed within this time frame.
  • Page 29: Speed Sintering

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.4 Operating modes Sintering After the holding time, the furnace opens at 1,580 °C (2,876 °F) and then cools down to 1,000 °C (1,832 °F) by itself. 1 minute before the door opens, the furnace displays a "countdown" of 60 seconds.
  • Page 30: Classic Sintering

    6 Operation Dentsply Sirona 6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed 6.4.3 Classic sintering (program starts at room temperature, no pre-heating required) Classic sintering refers to the conventional long-term sintering process, which does not require the furnace to be pre-heated. Stacking a maximum of two sintering trays is permitted.
  • Page 31: Use Of Argon In The Infire Htc Speed With Gas

    Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional) 6.5.1.2 Use of argon in the inFire HTC speed with gas management module With a factory-installed NPM function, the inFire HTC speed has an internal gas management module which requires an inlet pressure of 7-8 bar in order to provide the required volume flow of 1.1 l / min for NPM...
  • Page 32: Scope Of Supply

    ● Zirconium oxide beads, Ø 1.25 mm (200 g) for inCoris CC Retrofitting a gas management module 6.5.2.1 If you own a inFire HTC speed without an internal gas management module, you have the option of installing this later. 6.5.3 Enhanced program list The inFire HTC speed with NPM function has an enhanced program list.
  • Page 33: Sintering Pre-Sintered Npms (Cocr-Base)

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional) Determining the filling level of the compressed gas cylinder: The filling level of the compressed gas cylinder can be ascertained from the manometer. A new argon compressed gas cylinder is filled with 200 bar.
  • Page 34: Dentsply Sirona

    6 Operation Dentsply Sirona 6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed 1. Fill the sintering tray with zirconium oxide beads. The filling level should be between 1.5 and 2 cm. In this process, only use the zirconium oxide beads for NPM sintering (REF 64 13 640).
  • Page 35: Loading The Furnace With The Incoris Cc Set, From Serial Number

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional) 6. Remove the sintering tray after the furnace has cooled down. Remove the sintering tray using commercial crucible and ➢ investment tongs. Tip: From serial number 5000, the sintering bell can be removed using the supplied crucible fork.
  • Page 36: Dentsply Sirona

    6 Operation Dentsply Sirona 6.5 Non-precious metal (NPM) sintering (optional) Operating Instructions inFire HTC speed 3. Cover the sintering tray with the sintering bell. 4. Following the completed sintering process, the inCoris CC sintering set can be removed from the furnace, using the crucible fork, and allowed to cool outside the furnace.
  • Page 37: Replacing The Furnace Door

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.5 Non-precious metal (NPM) sintering (optional) 6.5.7 Replacing the furnace door (only for devices with serial numbers < 5000) WARNING Risk of burning The furnace door may be hot. ➢ Check the temperature of the furnace on the display.
  • Page 38: Programs

    The sintering program will not start. Programs Explanation With the inFire HTC speed sintering furnace, sintering is controlled via callable programs [ → 39]. ● inFire HTC speed has fixed programs for sintering certain dental oxide ceramics.
  • Page 39: Fixed Programs

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.6 Programs 6.6.2 Fixed programs Explanation inFire HTC speed features fixed programs for various materials and manufacturers. Overview Program Designation Sirona inCoris TZI/ZI super speed single Sirona inCoris TZI/ZI speed Sirona inCoris TZI/ZI speed + dry...
  • Page 40: Starting The Program

    6 Operation Dentsply Sirona 6.6 Programs Operating Instructions inFire HTC speed 6.6.4 Starting the program ➢ Press the START/STOP key to start the process.  The LCD displays shows the stand-by screen. In the top line, the status READY changes to RUN. Blinking or illumination of the stage LCD indicates that a process is running.
  • Page 41 Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.6 Programs Programming steps S2 to S4 ➢ Follow the same procedure as described for step S1 to program steps S2 to S4 (S2 key for step S2, etc.). In order to create a freely programmable fast sintering program, you must set a heating velocity of at least >...
  • Page 42: Entering The Program Values For Fast Sintering

    6 Operation Dentsply Sirona 6.6 Programs Operating Instructions inFire HTC speed 3. Press the S2 key (=YES) to confirm the loading procedure. Programming step S3 1. Press the S3 key. The cursor will blink in the "Heating velocity" column of line S3.
  • Page 43: Storing Program Values

    Dentsply Sirona 6 Operation Operating Instructions inFire HTC speed 6.6 Programs 6.6.6.4 Storing program values Explanation This section describes the procedure for storing changed program values of freely programmable settings (20-30). ✔ Call method Once program values have been entered with the keys S1 to S4, they can be saved.
  • Page 44: Default Settings

    7 Default settings Dentsply Sirona 7.1 Display and acoustic signal Operating Instructions inFire HTC speed Default settings Display and acoustic signal Explanation You can configure the display and acoustic signal via two dialog boxes. Call method Press function key "F". Dialog box 1 Key assignment ●...
  • Page 45: Adjustment Of The Start Time (Casting Time Function)

    Dentsply Sirona 7 Default settings Operating Instructions inFire HTC speed 7.2 Adjustment of the start time (casting time function) Adjustment of the start time (casting time function) Explanation When calculating the start time (if the casting-time function is used), the furnace controller assumes an average voltage of 230 V.
  • Page 46: Description Of Special Functions

    8 Description of special functions Dentsply Sirona 8.1 Sintering with ventilation (sintering with a small air gap) Operating Instructions inFire HTC speed Description of special functions Sintering with ventilation (sintering with a small air gap) This function means that the furnace door remains open with a small air gap at the first program level (S1) during the heat treatment process.
  • Page 47: Regular Care

    Dentsply Sirona 9 Regular care Operating Instructions inFire HTC speed 9.1 General care instructions Regular care General care instructions The high-temperature furnace does not require any special care. The casing may be cleaned with a mild detergent. NOTICE Damage to the heating unit Avoid all contamination in the heating room to prevent damage to the heating unit.
  • Page 48: Maintenance

    10 Maintenance Dentsply Sirona 10.1 Changing fuses F1/F2 Operating Instructions inFire HTC speed Maintenance 10.1 Changing fuses F1/F2 Replacement fuse The order number for the replacement fuse is: 61 32 448 (10 items). Required tools 1x (flat) screwdriver Position Fuses F1 and F2 are located at the bottom left on the rear side.
  • Page 49: Replacing The Battery (Only For Devices With Serial Numbers < 5000)

    Dentsply Sirona 10 Maintenance Operating Instructions inFire HTC speed 10.2 Replacing the battery (only for devices with serial numbers < 5000) 10.2 Replacing the battery (only for devices with serial numbers < 5000) The battery has a limited service life. The battery must be replaced 24 months after commissioning (date of battery installation).
  • Page 50: Malfunctions

    11 Malfunctions Dentsply Sirona 11.1 Error messages from electronic components Operating Instructions inFire HTC speed Malfunctions 11.1 Error messages from electronic components Error messages are displayed on the LCD screen. They are accompanied by an acoustic signal. Error message Cause Elimination DEFECTIVE SENSOR Thermocouple not functioning Call customer service.
  • Page 51: Miscellaneous Errors And Their Causes

    Dentsply Sirona 11 Malfunctions Operating Instructions inFire HTC speed 11.2 Miscellaneous errors and their causes 11.2 Miscellaneous errors and their causes Error Cause Elimination Wrong time displayed. Incorrect time in controller. Correct time settings (see function key "F" [ → 44]) Furnace does not heat up although Power was out for a long time during Check the power connection.
  • Page 52: Disposal

    12 Disposal Dentsply Sirona Operating Instructions inFire HTC speed Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
  • Page 53: Index

    Dentsply Sirona Index Operating Instructions inFire HTC speed Index Acoustic signal, 44, 44 Language, 44 Lift key, 18 Load program key, 18 Casting time key, 18 Casting-time function, 40, 45 CE mark, 12 Memory key, 18 Celsius, 44 Controller, 17...
  • Page 54: Dentsply Sirona

    Index Dentsply Sirona Operating Instructions inFire HTC speed Temperature, 19 Temperature scale, 44 Temperature steps, 19 The correction program, 45 Time display, 44 Ventilation slots, 14 61 30 459 D 3497 D 3497.201.01.19.02 09.2017...
  • Page 56 We reserve the right to make any alterations which may be required due to technical improvements. © Sirona Dental Systems GmbH Sprache: englisch Printed in Germany D 3497.201.01.19.02 09.2017 Ä.-Nr.: 124 274 Sirona Dental Systems GmbH Contact in the USA: Sirona Dental, Inc.

Table of Contents