Readings on the digital display
After power-on of the system, running dots initially appear on the digital
display for a brief time.
Then exposure program number P1, the maximum exposure time for
kV
mA
S
this program in seconds s and the "kV/mA " combination stored for this
exposure program are displayed.
If the exposure program number and a help message H... alternately
appear on the digital display, the help message must be processed first,
kV
mA
S
see "Help messages"  [→ 95]. The device is ready for operation when
no more help messages are displayed.
kV
mA
S
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it
to the starting position.
If the Ready LED above the R key starts flashing and error message H
301 appears on the display, briefly press return key R to bring the
rotating element into the starting position. The Ready LED then
switches off and the help message disappears.
The system is now ready to operate.
CAUTION
5 Operation