Error Description; Ex - Error Type - Dentsply Sirona Orthophos E Ceph Operating Instructions Manual

Hide thumbs Also See for Orthophos E Ceph:
Table of Contents

Advertisement

7 Malfunctions

7.3 Error description

98
Error description
7.3
7.3.1
Ex – Error type
NOTE
The unit must not be switched on/off constantly.
Constant switching on and off reduces the service life of individual
unit components and results in increased power consumption.
➢ After switching the unit off, wait for approx. 60 seconds before
switching it on again.
E1 – System warning/message
The error is in an acceptable tolerance range. Device operation is not
directly impaired.
1. Acknowledge the error message.
2. Contact your Customer Service.
Ä Continued device operation is ensured.
E2 – Overload
The error can be traced back to temporary overheating or something
similar.
1. Acknowledge the error message.
2. Wait for a moment and repeat the procedure step. If the error
reappears, extend the waiting time.
Ä The error no longer occurs after a certain waiting period.
3. If the error persists, contact your Customer Service.
E3 – Key pressed during power-up
The error results from an invalid signal state due to pushing buttons and
security signals during power-up.
1. Switch the unit off and on again. NOTE!  Observe waiting period!
2. If the error persists, contact your Customer Service.
E4 – Unassigned
E5 – Malfunction during exposure or exposure preparation
Error resulting from a certain system action triggered by the user which
could not be performed because a required (internal) partial function
(software or hardware) is not ready or fails.
1. Acknowledge the error message.
2. Repeat the last procedure step or exposure.
Ä The error no longer occurs.
3. If the error persists, contact your Customer Service.
Dentsply Sirona
Operating Instructions Orthophos E
67 11 001 D3352
D3352.201.12.01.02    03.2019

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos e

Table of Contents