Dentsply Sirona Orthophos E Ceph Operating Instructions Manual page 81

Hide thumbs Also See for Orthophos E Ceph:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Orthophos E
67 11 001 D3352
D3352.201.12.01.02    03.2019
The exposure memory of the unit is cleared when the unit is switched
off.
Images that have not been transferred to SIDEXIS are irretrievably
lost. The exposure must be repeated.
➢ Wait until the exposure data have be completely transferred. Do
not switch the unit off before the X-ray exposure is displayed on
the SIDEXIS screen.
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it
to the starting position.
IMPORTANT
Premature release of a new exposure is prevented by the automatic
exposure blocking function. This function is used for thermal
23
protection of the X-ray tube.
After the release button is pressed, the expiring cooling time in
seconds appears on the digital display. If you let go of the release
button before the cooling time has expired, the Ready LED above the
"R" key also starts flashing. After you press the "R" key, the program
data again appears on the digital display.
Another exposure can be released only after the cooling period has
elapsed.
ü The help message H... must not appear alternating on the digital
display of the Multipad.
Tip: If the doors of the X-ray room are not properly shut, the
message "H321" (Close the door) appears on the digital display of
the Multipad and on the remote control.
ü The exposure program, radiation time, and kV/mA values are
selected and appear on the digital display C.
ü The patient is positioned in the unit.
IMPORTANT
The patient's arms must hang down freely at his/her sides, He/she
should not tense or hunch his/her shoulders.
Advise the patient not to move his/her head in any way during the
exposure and check to make sure that this does not happen.
CAUTION
CAUTION
5 Operation
5.1 Acquiring the X-ray image
81