Dentsply Sirona Orthophos E Ceph Operating Instructions Manual page 19

Hide thumbs Also See for Orthophos E Ceph:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions Orthophos E
67 11 001 D3352
D3352.201.12.01.02    03.2019
Marking of focal spot:
Source-skin distance
Automatic exposure block-
ing:
Equipment class:
Class I device
Degree of protection
against electric shock:
Degree of protection
against ingress of water:
Year of manufacture:
Operating mode:
Long-term power output:
Anode material:
Exposure parameters for
determining leakage radia-
tion:
X-ray tube
Siemens SR 90/15 FN or
CEI OCX 100
3.2 Technical data
> 200 mm (8")
The duration of automatic exposure
blocking (cooling period) depends on
the set kV/mA level and the actual ex-
posure time. Depending on the tube
load, interval times of 8 s to 300 s are
automatically set by the system.
Example: For program P1 with expo-
sure data of 80 kV/14 mA and a radia-
tion time of 14.1 s, the pause duration
is 150 s.
IPX0
Type B device
Ordinary equipment (without protection
against ingress of water)
20XX
(on the rating plate)
Continuous operation
100 W
Tungsten
2  m A / 90 kV
3 Unit description
19