Error message structure

7.2
The error messages are displayed on the device in the form of an error
code. The display does not show any plain text error output.
The error codes are structured according to the following pattern: Ex yy
zz.
Explanation of abbreviations:
Ex – Error type
The x character provides a foundation for making quick decisions as to
how serious the error is and how to handle the error.
yy – Locality
Describes the impaired function of the device.
zz – Identification
Further specification of the error with a consecutive number.
7 Malfunctions
7.2 Error message structure