Dentsply Sirona
Operating Instructions Orthophos E
Program
Max. expo-
sure time
C1
9.1s
C2
9.1s
C3
9.4s
C3F 30x23
14.9s
C4
9.1s
67 11 001 D3352
D3352.201.12.01.02 03.2019
8.3.2
Dose area product parameters for Ceph-images
The ceph exposure values were measured with a semiconductor
detector, as very low dose values are hard to measure in a CT
ionization chamber.
Specification of the dose area product (DAP/absorbed dose) for lateral,
a.p., p.a. and carpus views
Factory-programmed values
DAP
2
mGycm
kV/mA
kV/mA
80/14
24
80/14
80/14
24
80/14
73/15
22
73/15
73/15
35
73/15
64/16
16
64/16
DAP
DAP
2
mGycm
mGycm
kV/mA
24
84/13
25
24
84/13
25
22
77/14
23
35
77/14
37
16
64/16
16
8 Program values
8.3 Dose information
DAP
2
2
mGycm
kV/mA
90/12
26
90/12
26
84/13
26
84/13
40
64/16
16
107