6 Service
6.1 Cleaning and care
92
6.1.3

Sterilization

WARNING
Infections can be transmitted from patient to patient.
Accessories that are not sterilized correctly can cause illness in
patients.
➢ All accessories that are suitable for sterilization should only be
sterilized in an autoclave at 132 °C (270 °F), with at least
4 minutes holding time and at 2.1 bar (30.5 psi) overpressure.
The following accessories can be sterilized:
In addition, also use the hygienic protective sleeves; see 'Hygienic
protective sleeves'  [→ 31].
WARNING
The hygienic protective sleeves are single use devices.
Contaminated hygienic protective sleeves can cause illness in
patients.
➢ Replace the hygienic protective sleeves after each patient.
Dentsply Sirona
Operating Instructions Orthophos E
13 °C
2
27 °F
67 11 001 D3352
D3352.201.12.01.02    03.2019