Inspection and maintenance

6.2
Inspection and preventive maintenance must be performed at
scheduled intervals to protect the health and safety of patients, users,
and other persons.
The information provided in the document Inspection and maintenance
and safety-related checks REF 67 10 961 should be helpful here. The
document can be downloaded at
http://www.dentsplysirona.com/manuals.
Annual inspection
In order to ensure the operational safety and functional reliability of your
product, you as the system owner should check the equipment at
regular intervals (at least once a year) or commission your dental depot
to do so.
In the event of visible external damage, commission your dental depot
with an inspection of the unit.
Maintenance by the service engineer
In addition to the annual check to be carried out by the system owner or
authorized persons, preventive maintenance must be performed after 4,
7 and 10 years, and then at two-year intervals.
Checking image quality
The image quality should be assessed by the system owner at regular
intervals, at least once a year.
On digital image receptor systems, the degree of postprocessing
(brightness or contrast adjustment) that is required in the image
processing software (e.g. SIDEXIS) to produce satisfactory results is
used as an assessment criterion.
If a unit defect cannot be ruled out in the event of an image that cannot
be evaluated, it is imperative a quality inspection is performed by an
authorized service technician.
6 Service
6.2 Inspection and maintenance