Rigol MSO1000Z Series Programming Manual page 38

Digital oscilloscope
Hide thumbs Also See for MSO1000Z Series:
Table of Contents

Advertisement

RIGOL
:CURSor:MANual:BXValue?
Syntax
:CURSor:MANual:BXValue?
Description
Query the X value of cursor B in the manual cursor measurement mode. The unit depends
on the horizontal unit currently selected.
Return
The query returns the X value of cursor B in scientific notation.
Format
Example
:CURSor:MANual:BXValue?
Related
:CURSor:MANual:BX
Commands
:CURSor:MANual:TUNit
:CURSor:MANual:BYValue?
Syntax
:CURSor:MANual:BYValue?
Description
Query the Y value of cursor B in the manual cursor measurement mode. The unit depends
on the vertical unit currently selected.
Return
When the signal source is CHANnel1|CHANnel2|CHANnel3|CHANnel4|MATH, the
Format
query returns the Y value of cursor B in scientific notation.
When the signal source is LA, the query returns the decimal value corresponding to
the sum of the binary weights of D15 bit to D0 bit at cursor B (the value of the
channel that is not turned on is 0 by default) .
Example
:CURSor:MANual:BYValue?
Related
:CURSor:MANual:BY
Commands
:CURSor:MANual:VUNit
:CURSor:MANual:XDELta?
Syntax
:CURSor:MANual:XDELta?
Description
Query the difference between the X values of cursor A and cursor B (BX-AX) in the manual
cursor measurement mode. The unit depends on the horizontal unit currently selected.
Return
The query returns the difference in scientific notation.
Format
Example
:CURSor:MANual:XDELta?
Related
:CURSor:MANual:AX
Commands
:CURSor:MANual:BX
:CURSor:MANual:TUNit
2-22
/*The query returns 4.000000e-06*/
/*The query returns -2.000000e+00*/
/*The query returns 8.000000e-06*/
MSO1000Z/DS1000Z Programming Guide
Chapter 2 Command System