Rigol MSO1000Z Series Programming Manual page 203

Digital oscilloscope
Hide thumbs Also See for MSO1000Z Series:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:RUNT:WLOWer
Syntax
:TRIGger:RUNT:WLOWer <NR3>
:TRIGger:RUNT:WLOWer?
Description
Set or query the pulse width lower limit in runt trigger. The default unit is s.
Parameter
Name
<NR3>
Explanation
This command is only available when the qualifier (refer to :TRIGger:RUNT:WHEN) is
GREater or GLESs.
When the qualifier is GLESs, the lower limit of the pulse width must be smaller than
the upper limit of the pulse width (:TRIGger:RUNT:WUPPer).
Return
The query returns the pulse width lower limit in scientific notation.
Format
Example
:TRIGger:RUNT:WLOWer 0.01
:TRIGger:RUNT:WLOWer?
:TRIGger:RUNT:ALEVel
Syntax
:TRIGger:RUNT:ALEVel <level>
:TRIGger:RUNT:ALEVel?
Description
Set or query the trigger level upper limit in runt trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
The query returns the trigger level upper limit in scientific notation.
Format
Example
:TRIGger:RUNT:ALEVel 0.16
:TRIGger:RUNT:ALEVel?
MSO1000Z/DS1000Z Programming Guide
Type
Range
When the qualifier is GREater, the range is from 8ns to
Real
10s; when the qualifier is GLESs, the range is from 8ns
to 9.99s.
/*Set the pulse width lower limit to 1ms*/
/*The query returns 1.000000e-03*/
Type
Range
(-5 x VerticalScale - OFFSet) to
Real
(5 x VerticalScale - OFFSet)
/*Set the trigger level upper limit to 160mV*/
/*The query returns 1.600000e-01*/
RIGOL
Default
1μs
Default
2V
2-187