Rigol MSO1000Z Series Programming Manual page 210

Digital oscilloscope
Hide thumbs Also See for MSO1000Z Series:
Table of Contents

Advertisement

RIGOL
:TRIGger:DELay:TYPe
Syntax
:TRIGger:DELay:TYPe <type>
:TRIGger:DELay:TYPe?
Description
Set or query the delay type in delay trigger.
Parameter
Name
<type>
Explanation
GREater: trigger when the time difference (△T) between the specified edges of tigger
source A and tigger source B is greater than the preset time limit.
LESS: trigger when the time difference (△T) between the specified edges of tigger
source A and tigger source B is lower than the preset time limit.
GLESs: trigger when the time difference (△T) between the specified edges of tigger
source A and tigger source B is greater than the lower limit of the preset time and
lower than the upper limit of the preset time. Note that the time lower limit must be
lower than the time upper limit.
GOUT: trigger when the time difference (△T) between the specified edges of tigger
source A and tigger source B is lower than the lower limit of the preset time or
greater than the upper limit of the preset time. Note that the time lower limit must be
lower than the time upper limit.
Return
The query returns GOUT, GRE, LESS, or GLES.
Format
Example
:TRIGger:DELay:TYPe GOUT
:TRIGger:DELay:TYPe?
Related
:TRIGger:DELay:TUPPer
Commands
:TRIGger:DELay:TLOWer
:TRIGger:DELay:TUPPer
Syntax
:TRIGger:DELay:TUPPer <NR3>
:TRIGger:DELay:TUPPer?
Description
Set or query the upper limit of the delay time in delay trigger. The default unit is s.
Parameter
Name
<NR3>
Explanation
This command is only available when the delay type (refer to :TRIGger:DELay:TYPe) is
LESS, GOUT, or GLESs.
Return
The query returns the upper limit of the delay time in scientific notation.
Format
Example
:TRIGger:DELay:TUPPer 0.002
:TRIGger:DELay:TUPPer?
Related
:TRIGger:DELay:TLOWer
Command
2-194
Type
Range
Discrete
{GREater|LESS|GLESs |GOUT}
/*Set the delay type to ><*/
/*The query returns GOUT*/
Type
Range
Real
16ns to 10s
/*Set the upper limit of the delay time to 2ms*/
/*The query returns 2.000000e-03*/
Chapter 2 Command System
MSO1000Z/DS1000Z Programming Guide
Default
GREater
Default
2μs