Rigol MSO1000Z Series Programming Manual page 36

Digital oscilloscope
Hide thumbs Also See for MSO1000Z Series:
Table of Contents

Advertisement

RIGOL
:CURSor:MANual:BX
Syntax
:CURSor:MANual:BX <x>
:CURSor:MANual:BX?
Description
Set or query the horizontal position of cursor B in the manual cursor measurement mode.
Parameter
Name
<x>
Explanation
The horizontal and vertical positions of the cursor are defined by the pixel coordinate of
the screen. The pixel coordinate of the screen ranges from (0,0) to (600,400). Wherein,
(0,0) is located at the left top corner of the screen and (600,400) is located at the right
bottom corner of the screen. The horizontal pixel range is from 0 to 600 and the vertical
pixel range is from 0 to 400.
Return
The query returns an integer between 5 and 594.
Format
Example
:CURSor:MANual:BX 200
:CURSor:MANual:BX?
:CURSor:MANual:AY
Syntax
:CURSor:MANual:AY <y>
:CURSor:MANual:AY?
Description
Set or query the vertical position of cursor A in the manual cursor measurement mode.
Parameter
Name
<y>
Explanation
The horizontal and vertical positions of the cursor are defined by the pixel coordinate
of the screen. The pixel coordinate of the screen ranges from (0,0) to (600,400).
Wherein, (0,0) is located at the left top corner of the screen and (600,400) is located
at the right bottom corner of the screen. The horizontal pixel range is from 0 to 600
and the vertical pixel range is from 0 to 400.
When the signal source of manual cursor measurement is LA, Y type cursor is not
required.
Return
The query returns an integer between 5 and 394.
Format
Example
:CURSor:MANual:AY 200
:CURSor:MANual:AY?
2-20
Type
Range
Integer
5 to 594
/*Set the horizontal position of cursor B to 200*/
/*The query returns 200*/
Type
Range
Integer
5 to 394
/*Set the vertical position of cursor A to 200*/
/*The query returns 200*/
Chapter 2 Command System
MSO1000Z/DS1000Z Programming Guide
Default
500
Default
100