RIGOL
:TRIGger:PATTern:LEVel
Syntax
:TRIGger:PATTern:LEVel <chan>,<level>
:TRIGger:PATTern:LEVel? <chan>
Description
Set or query the trigger level of the specified channel in pattern trigger. The unit is the
same as the current amplitude unit.
Parameter
Name
<chan>
<level>
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
This command is only available when the signal source is an analog channel.
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:PATTern:LEVel CHANnel2,0.16 /*Set the trigger level of CH2 to 160mV*/
:TRIGger:PATTern:LEVel? CHANnel2
2-178
Type
Range
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
(-5 x VerticalScale - OFFSet) to
Real
(5 x VerticalScale - OFFSet)
Chapter 2 Command System
/*The query returns 1.600000e-01*/
MSO1000Z/DS1000Z Programming Guide
Default
CHANnel1
0