Rigol MSO1000Z Series Programming Manual page 229

Digital oscilloscope
Hide thumbs Also See for MSO1000Z Series:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:SPI:CLEVel
Syntax
:TRIGger:SPI:CLEVel <level>
:TRIGger:SPI:CLEVel?
Description
Set or query the trigger level of the SCL channel in SPI trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
This command is only available when the signal source of the clock line is an analog
channel.
Return
The query returns the trigger level of the SCL channel in scientific notation.
Format
Example
:TRIGger:SPI:CLEVel 0.16
:TRIGger:SPI:CLEVel?
:TRIGger:SPI:DLEVel
Syntax
:TRIGger:SPI:DLEVel <level>
:TRIGger:SPI:DLEVel?
Description
Set or query the trigger level of the SDA channel in SPI trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
This command is only available when the signal source of the data line is an analog
channel.
Return
The query returns the trigger level of the SDA channel in scientific notation.
Format
Example
:TRIGger:SPI:DLEVel 0.16
:TRIGger:SPI:DLEVel?
MSO1000Z/DS1000Z Programming Guide
Type
Range
(-5 x VerticalScale - OFFSet) to
Real
(5 x VerticalScale - OFFSet)
/*Set the trigger level of the SCL channel to 160mV*/
/*The query returns 1.600000e-01*/
Type
Range
(-5 x VerticalScale - OFFSet) to
Real
(5 x VerticalScale - OFFSet)
/*Set the trigger level of the SDA channel to 160mV*/
/*The query returns 1.600000e-01*/
RIGOL
Default
0
Default
0
2-213

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents