Chapter 2 Command System
:TRIGger:DURATion:WHEN
Syntax
:TRIGger:DURATion:WHEN <when>
:TRIGger:DURATion:WHEN?
Description
Set or query the trigger condition in duration trigger.
Parameter
Name
<when>
Explanation
GREater: you need to specify a time (refer to :TRIGger:DURATion:TLOWer). The
oscilloscope triggers when the duration of the pattern is greater than the preset time.
LESS: you need to specify a time (refer to :TRIGger:DURATion:TUPPer). The
oscilloscope triggers when the duration of the pattern is lower than the preset time.
GLESs: you need to specify an upper limit of the time (refer
to :TRIGger:DURATion:TUPPer) and lower limit of the time (refer
to :TRIGger:DURATion:TLOWer). The oscilloscope triggers when the duration of the
pattern is lower than the preset upper limit of the time and greater than the preset
lower limit of the time.
Return
The query returns GRE, LESS, or GLES.
Format
Example
:TRIGger:DURATion:WHEN LESS
:TRIGger:DURATion:WHEN?
:TRIGger:DURATion:TUPPer
Syntax
:TRIGger:DURATion:TUPPer <NR3>
:TRIGger:DURATion:TUPPer?
Description
Set or query the duration time upper limit in duration trigger. The default unit is s.
Parameter
Name
<NR3>
Explanation
This command is available when the trigger condition (:TRIGger:DURATion:WHEN) is
LESS or GLESs.
Return
The query returns the duration time upper limit in scientific notation.
Format
Example
:TRIGger:DURATion:TUPPer 0.000003
:TRIGger:DURATion:TUPPer?
Related
:TRIGger:DURATion:TLOWer
Command
MSO1000Z/DS1000Z Programming Guide
Type
Range
Discrete
{GREater|LESS|GLESs}
Type
Range
Related to the trigger condition
Real
LESS: 8ns to 10s
GLESs: 16ns to 10s
/*Set the trigger condition to <*/
/*The query returns LESS*/
/*Set the duration time upper limit to 3μs*/
/*The query returns 3.000000e-06*/
RIGOL
Default
GREater
Default
2μs
2-181