Rigol MSO1000Z Series Programming Manual page 225

Digital oscilloscope
Hide thumbs Also See for MSO1000Z Series:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:IIC:CLEVel
Syntax
:TRIGger:IIC:CLEVel <level>
:TRIGger:IIC:CLEVel?
Description
Set or query the trigger level of SCL in I2C trigger. The unit is the same as the
current amplitude unit.
Parameter
Explanation
Return Format
The query returns the trigger level of SCL in scientific notation.
Example
:TRIGger:IIC:CLEVel 0.16
:TRIGger:IIC:CLEVel?
:TRIGger:IIC:DLEVel
Syntax
:TRIGger:IIC:DLEVel <level>
:TRIGger:IIC:DLEVel?
Description
Set or query the trigger level of SDA in I2C trigger. The unit is the same as the
current amplitude unit.
Parameter
Explanation
Return Format
The query returns the trigger level of SDA in scientific notation.
Example
:TRIGger:IIC:DLEVel 0.16
:TRIGger:IIC:DLEVel?
MSO1000Z/DS1000Z Programming Guide
Name
Type
Range
(-5 x VerticalScale - OFFSet) to
<level>
Real
(5 x VerticalScale - OFFSet)
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
This command is only available when the signal source of the clock line is an
analog channel.
Name
Type
Range
(-5 x VerticalScale - OFFSet) to
<level>
Real
(5 x VerticalScale - OFFSet)
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
This command is only available when the signal source of the data line is an
analog channel.
/*Set the trigger level of SCL to 160mV*/
/*The query returns 1.600000e-01*/
/*Set the trigger level of SDA to 160mV*/
/*The query returns 1.600000e-01*/
RIGOL
Default
0
Default
0