Chapter 2 Command System
:TRIGger:SLOPe:WINDow
Syntax
:TRIGger:SLOPe:WINDow <window>
:TRIGger:SLOPe:WINDow?
Description
Set or query the vertical window type in slope trigger.
Parameter
Name
<window>
Explanation
TA: only adjust the upper limit of the trigger level, refer to :TRIGger:SLOPe:ALEVel.
TB: only adjust the lower limit of the trigger level, refer to :TRIGger:SLOPe:BLEVel.
TAB: adjust the upper and lower limits of the trigger level at the same time, refer
to
Return
The query returns TA, TB, or TAB.
Format
Example
:TRIGger:SLOPe:WINDow TB
:TRIGger:SLOPe:WINDow?
:TRIGger:SLOPe:ALEVel
Syntax
:TRIGger:SLOPe:ALEVel <level>
:TRIGger:SLOPe:ALEVel?
Description
Set or query the upper limit of the trigger level in slope trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
The query returns the upper limit of the trigger level in scientific notation.
Format
Example
:TRIGger:SLOPe:ALEVel 0.16
:TRIGger:SLOPe:ALEVel?
Related
:TRIGger:SLOPe:BLEVel
Command
MSO1000Z/DS1000Z Programming Guide
Type
Range
Discrete
{TA|TB|TAB}
:TRIGger:SLOPe:ALEVel
Type
Range
(-5 x VerticalScale - OFFSet) to
Real
(5 x VerticalScale - OFFSet)
and :TRIGger:SLOPe:BLEVel.
/*Set the vertical window type to TB*/
/*The query returns TB*/
/*Set the upper limit of the trigger level to 160mV*/
/*The query returns 1.600000e-01*/
RIGOL
Default
TA
Default
2V