RIGOL
:TRIGger:DURATion:TLOWer
Syntax
:TRIGger:DURATion:TLOWer <NR3>
:TRIGger:DURATion:TLOWer?
Description
Set or query the duration time lower limit in duration trigger. The default unit is s.
Parameter
Name
<NR3>
Explanation
This command is available when the trigger condition (:TRIGger:DURATion:WHEN) is
GREater or GLESs.
Return
The query returns the duration time lower limit in scientific notation.
Format
Example
:TRIGger:DURATion:TLOWer 0.000003
:TRIGger:DURATion:TLOWer?
Related
:TRIGger:DURATion:TUPPer
Command
2-182
Type
Range
Real
8ns to 9.99s
Chapter 2 Command System
/*Set the duration time lower limit to 3μs*/
/*The query returns 3.000000e-06*/
MSO1000Z/DS1000Z Programming Guide
Default
1μs