Chapter 2 Command System
:TRIGger:EDGe:LEVel
Syntax
:TRIGger:EDGe:LEVel <level>
:TRIGger:EDGe:LEVel?
Description
Set or query the trigger level in edge trigger. The unit is the same as the current
amplitude unit of the signal source selected.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to the
the
This command is only available when the signal source is an analog channel.
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:EDGe:LEVel 0.16
:TRIGger:EDGe:LEVel?
MSO1000Z/DS1000Z Programming Guide
Type
Range
(-5 x VerticalScale - OFFSet) to
Real
(5 x VerticalScale - OFFSet)
:CHANnel<n>:OFFSet
:CHANnel<n>:SCALe
command.
/*Set the trigger level to 160mV*/
/*The query returns 1.600000e-01*/
Default
0
command. For OFFSet, refer to
RIGOL