Optional Configuration
8.13.2

Fault timeout

If a fault is detected, the transmitter always sets the "alarm active" status bit immediately. Fault
actions for the transmitter outputs and digital communications may be implemented immediately or
may be delayed until the fault timeout expires. During the fault timeout, outputs continue to report
their last measured value.
The default fault timeout value is
The fault timeout is not applicable to all faults. See Table 8-8 for information about which faults are
affected.
8.14

Configuring the display

• Model 1700 AN
• Model 1700 IS
• Model 2700 AN
• Model 2700 IS
• Model 2700 CIO
8.14.1

Update period

The update period (or display rate) parameter controls how often the display is refreshed with current
data. The default is 0.2 seconds. The range is 0.10 seconds to 10 seconds. The Update Period value
applies to all process variables.
8.14.2

Language

The display can be configured to use any of the following languages for data and menus:
English
French
German
Spanish
8.14.3

Enabling and disabling display functions

Table 8-9 lists the display functions and describes their behavior when enabled or disabled.
Table 8-9
Display parameters
Parameter
Enabled
(1)(2)
Totalizer start/stop
Operators can start or stop totalizers using the
display.
(1)
Totalizer reset
Operators can reset the mass and volume
totalizers.
Auto scroll
The display automatically scrolls through each
process variable at a configurable rate.
Off-line menu
Operators can access the off-line menu (zero,
simulation, and configuration).
(3)
Off-line password
Operators must enter the display password to
access the off-line menu.
Alarm menu
Operators can access the alarm menu
(viewing and acknowledging alarms).
Configuration and Use Manual
, meaning that fault actions will be implemented immediately.
0
If your transmitter has a display, you can enable or disable specific display
functions, specify the process variables to be shown on the display, and set a variety
of parameters that control display behavior.
Disabled
Operators cannot start or stop totalizers using
the display.
Operators cannot reset the mass and volume
totalizers.
Operators must
to view process
Scroll
variables.
Operators cannot access the off-line menu.
Operators can access the off-line menu
without the display password.
Operators cannot access the alarm menu.
85