Measurement Performance
Detailed test data with the Communicator
Note: Requires Smart Meter Verification. No detailed test data is available with the original version
of the meter verification application.
For each Smart Meter Verification test, the following data is stored on the transmitter:
Powered-on hours at the time of the test
Test result
Stiffness of the left and right pickoffs, shown as percentage variation from the factory value. If
the test aborted, 0 is stored for these values.
Abort code, if applicable
The Communicator also provides a trend function that allows you to compare the results of the 20
tests, viewed as either a table or a graph.
To view this data, see Figure 10-10.