Measurement Performance
Figure 10-10
Meter verification test data – Communicator
Online >
1 Overview >
3 Shortcuts >
6 Meter Verification
10.3.5

Setting up automatic or remote execution of the meter verification test

Note: Requires Smart Meter Verification. Scheduling is not available with the original version of the
meter verification application.
There are three ways to execute a Smart Meter Verification test automatically:
Define it as an event action (using the dual-setpoint event model)
Set up a one-time automatic execution
Set up a recurring execution
Configuration and Use Manual
1 Run Meter Verification
2 View Test Results
3 Schedule Meter Verification
2
1 Run Counter
2 Running Hours
3 Test Result
4 Abort Code
5 LPO Stiffness
5 RPO Stiffness
7 Show Results Table
8 Show Results Plot
7
Test Result #x
Run Counter
Running Hours
Test Result
Abort Code
LPO Stiffness
RPO Stiffness
Abort
Online >
3 Service Tools >
4 Maintenance >
1 Routine Maintenance >
3 Meter Verification
OK
To previous test
115