Measurement Performance
Figure 10-17
Temperature calibration – ProLink II
Temperature Offset calibration
124
Fill sensor with low-
temperature fluid
Wait until sensor achieves
thermal equilibrium
ProLink Menu >
Calibration >
Temp offset cal
Enter temperature of low-
temperature fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Temperature Slope calibration
Fill sensor with high-
temperature fluid
Wait until sensor achieves
thermal equilibrium
ProLink Menu >
Calibration >
Temp slope cal
Enter temperature of high-
temperature fluid
Do Cal
Calibration in Progress
light turns red
Calibration in Progress
light turns green
Close
Done
Micro Motion Series 1000 and Series 2000 Transmitters