Hart/Bell 202 Connections - Emerson Micro Motion 1000 Series Configuration And Use Manual

Hide thumbs Also See for Micro Motion 1000 Series:
Table of Contents

Advertisement

Connecting with ProLink II or Pocket ProLink Software
3.5.2

HART/Bell 202 connections

Connecting a HART device to the transmitter's primary mA output terminals
could cause transmitter output error.
If the primary mA output is being used for flow control, connecting a HART device
to the output loop could cause the transmitter's 4–20 mA output to change, which
would affect flow control devices.
Set control devices for manual operation before connecting a HART device to the
transmitter's primary mA output loop.
Follow the instructions below to make the connection.
1. Connect the HART interface to your PC's serial or USB port. Then connect the leads of the
HART interface to terminals 21 and 22 on the transmitter (see Figure 3-9).
Figure 3-9
HART/Bell 202 connections to Model 1500 or 2500
250–600 Ω
2. Add 250–600 Ω resistance to the connection, as required.
3. Start ProLink II. Choose
4. In the screen that appears, set
are automatically set to the values required by HART protocol. Specify the remaining
connection parameters as shown in Table 3-3.
Table 3-3
HART connection parameters for ProLink II
Connection parameter
Address/Tag
COM port
26
CAUTION
Connection > Connect to Device
to HART Bell 202.
Protocol
HART setting
Configured HART polling address (default = 0)
COM port assigned to PC serial port
Micro Motion Series 1000 and Series 2000 Transmitters
.
,
, and
Baud rate
Stop bits
Parity

Advertisement

Table of Contents
loading

Table of Contents