5.5.1

Configure Fault Timeout

Display
ProLink II
ProLink III
Field Communicator Configure > Alert Setup > Alert Severity > Fault Timeout
Overview
Fault Timeout controls the delay before fault actions are performed.
Restriction
Fault Timeout is applied only to the following alarms (listed by Status Alarm Code): A003, A004, A005,
A008, A016, A017, A033. For all other alarms, fault actions are performed as soon as the alarm is
detected.
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alarm condition is
detected.
The fault timeout period begins when the transmitter detects an alarm condition. During
the fault timeout period, the transmitter continues to report its last valid measurements.
If the fault timeout period expires while the alarm is still active, the fault actions are
performed. If the alarm condition clears before the fault timeout expires, no fault actions
are performed.
Tip
ProLink II allows you to set Fault Timeout in two locations. However, there is only one parameter, and
the same setting is applied to all outputs.
5.5.2

Configure Status Alarm Severity

Display
ProLink II
ProLink III
Field Communicator Configure > Alert Setup > Alert Severity > Set Alert Severity
90
Not available
ProLink > Configuration > Analog Output > Last Measured Value Timeout
ProLink > Configuration > Frequency/Discrete Output > Frequency > Last Measured Value Timeout
Device Tools > Configuration > Fault Processing
Not available
ProLink > Configuration > Alarm > Severity
Device Tools > Configuration > Alert Severity
Micro Motion
®
Model 2700 Transmitters with Analog Outputs

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents