Emerson Micro Motion 1000 Series Configuration And Use Manual page 93

Hide thumbs Also See for Micro Motion 1000 Series:
Table of Contents

Advertisement

Optional Configuration
Table 8-8
Status alarms and severity levels continued
Communicator message
Alarm code
ProLink II message
(1)
A024
(E)EPROM Program Corrupt (Core Processor)
(E)EPROM Program Corrupt (CP)
(1)
A025
Protected Boot Sector Fault
Protected Boot Sector Fault (CP)
A026
Sensor/Xmtr Communication Error
Sensor/Transmitter Communication Error
A027
Security Breach
Security Breach
A028
Sensor/Xmtr Communication Failure
Core Processor Write Failure
(2)
A031
Undefined
Low Power
(3)
A032
Meter Verification Fault Alarm
Meter Verification/Outputs In Fault
(4)
A032
Outputs Fixed during Meter Verification
Meter Verification In Progress and Outputs
Fixed
(2)
A033
Sensor OK / Tubes Stopped by Process
Sensor OK/Tubes Stopped by Process
(4)
A034
Meter Verification Failed
Meter Verification Failed
(4)
A035
Meter Verification Aborted
Meter Verification Aborted
A100
Primary mA Output Saturated
Primary mA Output Saturated
A101
Primary mA Output Fixed
Primary mA Output Fixed
A102
Drive Over-Range / Partially Full Tube
Drive Overrange
(1)
A103
Data Loss Possible
Data Loss Possible (Tot and Inv)
A104
Calibration-In-Progress
Calibration in Progress
A105
Slug Flow
Slug Flow
A106
Burst Mode Enabled
Burst Mode Enabled
A107
Power Reset Occurred
Power Reset Occurred
Configuration and Use Manual
Default
severity
Configurable
Fault
No
Fault
No
Fault
No
Fault
No
Fault
No
Fault
No
Fault
No
(5)
Varies
No
Fault
Yes
Info
Yes
Info
Yes
(6)
Info
Yes
(6)
Info
Yes
Info
Yes
Info
Yes
(6)
Info
Yes
Info
Yes
(6)
Info
Yes
Info
Yes
Affected by
fault timeout
No
No
No
No
No
No
No
No
Yes
No
No
No
No
No
No
No
No
No
No
83