Pressure Compensation, Temperature Compensation, and Polling
Note: If your core processor is v2.1 or earlier, and you configure the transmitter for external
temperature compensation, the temperature value from the compensation procedure will replace the
sensor value in all calculations that require temperature data. If your core processor is v2.2 or later,
the temperature value from the compensation procedure is used only for enhanced density and
petroleum measurement calculations.
Note: If you configure a static temperature value, ensure that it is accurate. If you configure polling
for temperature, ensure that the external temperature measurement device is accurate and reliable.
9.3.1
Configuration
To configure external temperature compensation:
With ProLink II, see Figure 9-3
With the Communicator, see Figure 9-4
Figure 9-3
Configuring external temperature compensation with ProLink II
Enable
View Menu >
Preferences
Enable Use External
Temperature
Apply
(1) Setting the temperature measurement unit is
optional.
(2) See Section 9.4.
(3) If previously configured. Polling for pressure is
allowed. See Section 9.4.
92
Configure
Poll?
ProLink >
Configuration >
Temperature
(1)
Enter Temperature units
Apply
Go to Polling
(2)
Setup
Micro Motion Series 1000 and Series 2000 Transmitters
Static?
Disable polling for
(3)
temperature
ProLink >
Configuration >
Temperature
Enter External temperature
Apply
Done