4. Monitor Response

Configuring an LXI State Trigger

Task
Action
1. Place the
Various SCPI commands may be used.
instrument
in the state
to be recalled
2. Save the
:MMEM:STOR:STATe
state
"c:\States\myState.state"
3. Place the
Various SCPI commands may be used
instrument
in a new
state
4. Set up the
:LXI:EVENt:INPut:LAN:LOCation
state trigger
"LAN0","c:\States\myState.state"
5. Enable the
:LXI:EVENt:INPut:LAN:ENABled
state trigger
"LAN0",1
Recalling a state in response to a peer to peer message is done in a
similar way, although only a SCPI interface is available.
Note that the Input Event Location parameter is also a part of the
instrument's state. This means that each saved state may point to new
states to be recalled on subsequent peer to peer messages. For example,
in
"Configuring an LXI State
Location parameter for LAN1 to "c:\States\anotherState.state". After
Chapter 3
Follow
"Verifying LXI Output
LAN Events" on page
90.
Trigger", Task 1 may include setting the
Measurement Synchronization with LXI-B
Configuring LXI Triggers
This may be done
assuming
"Configuring
an LXI Trigger Using
SCPI" on page 90
"Configuring an LXI
Trigger Using the Front
Panel" on page 92
complete.
Notes
This step is up to the user's
discretion
The path must exist, and is up to
the user's discretion.
This step is up to the user's
discretion
Parameter one may be any Event
Name reported by the query
:LXI:EVENt:INPut:LAN:LIST?
Parameter 2 must match the
path used in Task 2.