Agilent Technologies X Series Programmer's Manual page 58

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Max
SCPI Status Bits/OPC
Dependencies
Questionable Frequency Positive Transition This command determines which bits in the
Questionable Frequency Condition register will set the corresponding bit in the
Questionable Frequency Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Register
Questionable Integrity Condition This query returns the decimal value of the sum of the
bits in the Questionable Integrity Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
Mode
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Enable This command determines which bits in the Questionable
Integrity Condition Register will set bits in the Questionable Integrity Event register,
which also sets the Integrity Summary bit (bit 9) in the Questionable Register. The
58
32767
Sequential command
All
:STATus:QUEStionable:FREQuency:PTRansition
<integer>
:STATus:QUEStionable:FREQuency:PTRansition?
STAT:QUES:FREQ:PTR 2 Frequency Reference 'became
unlocked' will be reported to the Frequency Summary of the
Status Questionable register.
32767
0
32767
Sequential command
All
:STATus:QUEStionable:INTegrity:CONDition?
STAT:QUES:INT:COND?
0
Sequential command