Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
The register requires that the associated PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Mode
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Questionable Temperature Negative Transition This command determines which bits in the
Questionable Temperature Condition register will set the corresponding bit in the
Questionable Temperature Event register when the condition register bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Temperature Positive Transition This command determines which bits in the
Questionable Temperature Condition register will set the corresponding bit in the
Questionable Temperature Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
68
All
:STATus:QUEStionable:TEMPerature[:EVENt]?
STAT:QUES:TEMP?
0
Sequential command
All
:STATus:QUEStionable:TEMPerature:NTRansition
<integer>
:STATus:QUEStionable:TEMPerature:NTRansition?
STAT:QUES:TEMP:NTR 1 Reference Oscillator Oven not cold
will be reported to the Temperature Summary of the Status
Questionable register.
0
0
32767
Sequential command
All