Agilent Technologies X Series Programmer's Manual page 59

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

variable <integer> is the sum of the decimal values of the bits you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Event Query This query returns the decimal value of the sum of the
bits in the Questionable Integrity Event register.
The register requires that the associated PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Mode
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Negative Transition This command determines which bits in the
Questionable Integrity Condition register will set the corresponding bit in the
Questionable Integrity Event register when the condition register bit has a negative
transition (1 to 0)
The variable <integer> is the sum of the decimal values of the bits that you want to enable.
Mode
Remote Command
Chapter 1
All
:STATus:QUEStionable:INTegrity:ENABle
:STATus:QUEStionable:INTegrity:ENABle?
STAT:QUES:INT:ENAB 8 Measurement Uncalibrated
Summary will be reported to the Integrity Summary of the
Status Questionable register.
32767
0
32767
Sequential command
All
:STATus:QUEStionable:INTegrity[:EVENt]?
STAT:QUES:INT?
0
Sequential command
All
:STATus:QUEStionable:INTegrity:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:NTRansition?
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
<integer>
59