Status Subsystem Commands And Queries - Agilent Technologies X Series Programmer's Manual

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

signal. All the bits are summary bits from lower-level event registers.
Bit
Condition
3
Power summary
4
Temperature summary
5
Frequency summary
8
Calibration summary
9
Integrity summary

:STATus Subsystem Commands and Queries

The :STATus subsystem controls the SCPI-defined instrument status reporting structures.
Each status register has a set of five commands used for querying or masking that
particular register.
Numeric values for bit patterns can be entered using decimal or hexadecimal
representations. (i.e. 0 to 32767 is equivalent to #H0 to #H7FFF. It is also equal to all ones,
111111111111111) See the SCPI Basics information about using bit patterns for variable
parameters.
Operation Register
Operation Condition Query This query returns the decimal value of the sum of the bits in
the Status Operation Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Operation Enable This command determines which bits in the Operation Event register,
will set the Operation Status Summary bit (bit 7) in the Status Byte Register. The variable
<integer> is the sum of the decimal values of the bits you want to enable.
Chapter 1
Operation
The instrument hardware has detected a power
unleveled condition.
The instrument is still warming up.
The instrument hardware has detected an unlocked
condition or a problem with the external frequency
reference.
The instrument has detected a hardware problem
while doing the automatic internal alignment
process.
The instrument has detected a questionable
measurement condition such as: bad timing, bad
signal/data, timeout problem, signal overload, or
"meas uncal".
All
:STATus:OPERation:CONDition?
STAT:OPER:COND?
0
Sequential command
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
43

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Exa n9010aMxa n9020a

Table of Contents