Agilent Technologies X Series Programmer's Manual page 102

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Measurement Synchronization with LXI-B
Setting Up and Executing a List Sweep Measurement
15. Choose the MXA's
PTP domain
16. Choose the MXA's
LXI domain
17. Turn off auto
alignments
18. Put the MXA into
SA mode.
19. Turn off continuous
measurements.
20. Put the MXA into
the list sweep
measurement
21. Send the MXA a list
of frequencies.
22. Configure the
MXA's list sweep
parameters.
23. Configure the MXA
to trigger on an alarm.
102
:LXI:CLOCk:PTP:DOMain 0
:LXI:EVENt:DOMain 0
CAL:AUTO:OFF
INST:SEL SA
INIT:CONT OFF
CONF:LIST
LIST:FREQ 100 MHz, 200 MHz, 300
MHz, 400 MHz, 500 MHz
LIST:ATT 10dB
LIST:BAND:RES:TYPE FLAT
LIST:BAND:RES 300 kHz
LIST:BAND:VID 3MHz
LIST:SWE:TIME 1e-6 s
LIST:TRIG:DEL 0
LIST:DET RMS
:LIST:TRIGger:SOURce ALARM
This command is sent
to the MXA.
The parameter value
should match the one
used in Task 3.
This command is sent
to the MXA.
The parameter value
should match the one
used in Task 4.
This command is sent
to the MXA.
Auto alignments will
introduce timing
variability.
This command is sent
to the MXA.
This command is sent
to the MXA.
This command is sent
to the MXA.
This command is sent
to the MXA.
These commands are
sent to the MXA.
This command is sent
to the MXA.
Chapter 3