Agilent Technologies X Series Programmer's Manual page 64

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Example
Preset
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Uncalibrated Negative Transition This command determines which
bits in the Questionable Integrity Uncalibrated Condition register will set the
corresponding bit in the Questionable Integrity Uncalibrated Event register when the
condition register bit has a negative transition (1 to 0). The variable <integer> is the sum
of the decimal values of the bits that you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Uncalibrated Positive Transition This command determines which bits
in the Questionable Integrity Uncalibrated Condition register will set the corresponding
bit in the Questionable Integrity Uncalibrated Event register when the condition register
bit has a positive transition (0 to 1). The variable <integer> is the sum of the decimal
values of the bits that you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
64
STAT:QUES:INT:UNC?
0
Sequential command
All
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition?
STAT:QUES:INT:UNC:NTR 1 Oversweep cleared will be reported to
the Integrity Summary of the Status Questionable register.
0
0
32767
Sequential command
All
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition?
STAT:QUES:INT:UNC:PTR 1 Oversweep (Meas Uncal) occurred will
be reported to the Integrity Summary of the Status Questionable
register.
32767
0
32767
Chapter 1