SCPI to get up and running.
Task
Action
1. Reset the
*RST
instrument
2. Set LXI Output
:LXI:EVENt:OUTPut:LAN:SOURce
LAN Event's
"LAN0","Sweeping"
source
3. Enable the LXI
:LXI:EVENt:LAN:OUTPut:ENABled
Output LAN
"LAN0",1
Event
4. Enable the
:LXI:EVENt:STATus:ENABled
Instrument Event
"Sweeping",1
5. Turn
:INIT:CONTinuous ON
continuous
measurement
mode on
Chapter 3
Measurement Synchronization with LXI-B
Peer to Peer Messaging
Notes
This puts the instrument in
its default state.
Hook the LAN0 LXI Output
LAN Event up to the
Sweeping instrument event.
Note that valid entries for the
first parameter are LAN0,
LAN1, LAN2, LAN3, LAN4,
LAN5, LAN6, LAN7
Note that valid entries for the
second parameter include the
list of Instrument Events in
Table 3 as well as the LXI
Input LAN Events
LAN0-LAN7. The instrument
will echo received LXI Input
LAN Events when used as a
source.
Turn the LAN0 LXI Output
LAN Event on.
Note that the parameter
should match the first
parameter in Task 2.
Turn the Sweeping
instrument event on.
Note that the parameter
should match the second
parameter in Task 2.