current conditions.
Mode
Remote
Command
Example
Preset
SCPI Status
Bits/OPC
Dependencies
Questionable Integrity Uncalibrated Enable This command determines which bits in the
Questionable Integrity Uncalibrated Condition Register will set bits in the Questionable
Integrity Uncalibrated Event register, which also sets the Data Uncalibrated Summary bit
(bit 3) in the Questionable Integrity Register. The variable <integer> is the sum of the
decimal values of the bits you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Uncalibrated Event Query This query returns the decimal value of
the sum of the bits in the Questionable Integrity Uncalibrated Event register.
The register requires that the associated PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Mode
Remote Command
Chapter 1
All
:STATus:QUEStionable:INTegrity:UNCalibrated:CONDition?
STAT:QUES:INT:UNC:COND?
0
Sequential command
All
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
:STATus:QUEStionable:INTegrity:UNCalibrated:ENABle
?
STAT:QUES:INT:UNC:ENAB 1 Oversweep (Meas Uncal) will
be reported to the Integrity Summary of the Status
Questionable register.
32767
0
32767
Sequential command
All
:STATus:QUEStionable:INTegrity:UNCalibrated[:EVENt]?
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
63