Agilent Technologies X Series Programmer's Manual page 60

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Positive Transition This command determines which bits in the
Questionable Integrity Condition register will set the corresponding bit in the
Questionable Integrity Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Integrity Signal Register
Questionable Integrity Signal Condition This query returns the decimal value of the sum of
the bits in the Questionable Integrity Signal Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
Mode
Remote Command
Example
Preset
60
STAT:QUES:INT:NTR 8 Measurement 'regained calibration'
Summary will be reported to the Integrity Summary of the
Status Questionable register.
0
0
32767
Sequential command
All
:STATus:QUEStionable:INTegrity:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:PTRansition?
STAT:QUES:INT:PTR 8 Measurement 'became uncalibrated'
Summary will be reported to the Integrity Summary of the
Status Questionable register.
32767
0
32767
Sequential command
All
:STATus:QUEStionable:INTegrity:SIGNal:CONDition?
STAT:QUES:INT:SIGN:COND?
0
Chapter 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Exa n9010aMxa n9020a

Table of Contents