Measurement Synchronization with LXI-B
Configuring LXI Triggers

Configuring an LXI Trigger Using the Front Panel

Task
1. Put the MXA into
SA or IQ Analyzer
mode.
2. Set the trigger
source to LAN
3. Select a Trigger
LAN Event
4. Enable the
Trigger LAN Event
5. Turn on
continuous
measurement
mode.

Verifying an LXI Trigger using LXI Interactive

Task
1. Open Interactive LXI
2. Choose LAN Event
3. Send LAN Event
92
Action
Press the hard key Mode followed by
the softkey Spectrum Analyzer or IQ
Analyzer (Basic)
Press the Trigger hardkey.
Press the softkeys More (1 of 2), LXI,
LAN.
Press the LAN softkey again, then
click on the event that should cause
the instrument to trigger.
Click the Enabled soft key.
Press the hard key labeled "Cont".
Action
From the Windows taskbar,
select Start, All Programs,
Agilent IO Libraries Suite,
Utilities, Interactive LXI
Select the desired LAN trigger
from the Event ID drop-down
menu.
Press the Send button
Notes
This is equivalent to Task 2
in
"Configuring an LXI
Trigger Using SCPI" on
page
90.
You will see a list of Trigger
LAN Event names.
This is equivalent to Task 3
in the procedure
"Configuring
an LXI Trigger Using SCPI"
on page
90.
Note that Tasks 3 & 4 may be
repeated for as many LAN
Events as desired.
Notes
This step should be
performed on a PC
connected to the same
subnet as the instrument.
This puts the trigger out
on the line.