Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
The preset condition is to have all bits in this enable register set to 0. To have any
Operation Events reported to the Status Byte Register, one or more bits need to be set to 1.
R&D: There is little reason to have any bits enabled for typical manufacturing tests.
Enabling bits in this register would be of more value during test development.
Mode
Remote Command
Example
Preset
Min
Max
SCPI Status Bits/OPC
Dependencies
Operation Event Query This query returns the decimal value of the sum of the bits in the
Operation Event register.
The register requires that the associated PTR or NTR filters be set before a condition
register bit can set a bit in the event register. The data in this register is latched until it is
queried. Once queried, the register is cleared.
Mode
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Operation Negative Transition This command determines which bits in the Operation
Condition register will set the corresponding bit in the Operation Event register when the
condition register bit has a negative transition (1 to 0). The variable <integer> is the sum
of the decimal values of the bits that you want to enable.
Mode
Remote Command
Example
44
All
:STATus:OPERation:ENABle <integer>
:STATus:OPERation:ENABle?
STAT:OPER:ENAB 1 Sets the register so that Align Now
operation is reported to the Status Byte Register.
0
0
32767
Sequential command
All
:STATus:OPERation[:EVENt]?
STAT:OPER?
0
Sequential command
All
:STATus:OPERation:NTRansition <integer>
:STATus:OPERation:NTRansition?
STAT:OPER:NTR 1 Align Now operation complete is reported
to the Status Byte Register.