Agilent Technologies X Series Programmer's Manual page 48

Signal analyzer
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Questionable Calibration Enable This command determines which bits in the Questionable
Calibration Condition Register will set bits in the Questionable Calibration Event register,
which also sets the Calibration Summary bit (bit 8) in the Questionable Register. The
variable <integer> is the sum of the decimal values of the bits you want to enable.
Mode
Remote Command
Example
Min
Max
SCPI Status Bits/OPC
Dependencies
Questionable Calibration Event Query This query returns the decimal value of the sum of
the bits in the Questionable Calibration Event register.
NOTE: The register requires that the associated PTR or NTR filters be set before a
condition register bit can set a bit in the event register. The data in this register is latched
until it is queried. Once queried, the register is cleared.
Mode
Remote Command
Example
Preset
SCPI Status Bits/OPC
Dependencies
Questionable Calibration Negative Transition This command determines which bits in the
Questionable Calibration Condition register will set the corresponding bit in the
Questionable Calibration Event register when the condition register bit has a negative
transition (1 to 0). The variable <integer> is the sum of the decimal values of the bits that
48
:STATus:QUEStionable:CALibration:CONDition?
STAT:QUES:CAL:COND?
0
Sequential command
All
:STATus:QUEStionable:CALibration:ENABle <integer>
:STATus:QUEStionable:CALibration:ENABle?
STAT:QUES:CAL:ENAB 16384 Can be used to query if an
alignment is needed, if you have turned off the automatic
alignment process.
0
32767
Sequential command
All
:STATus:QUEStionable:CALibration[:EVENt]?
STAT:QUES:CAL?
0
Sequential command
Chapter 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Exa n9010aMxa n9020a