Task
1. Put the MXA into SA or IQ
Analyzer mode.
2. Navigate to the System
panel.
3. Navigate to the LXI
Output LAN Event list.
4. Choose an LXI Output
LAN Event
5. Choose an instrument
event source.
7. Enable the LXI Output
LAN Event
8. Verify that the instrument
is in continuous
measurement mode.
9. Verify that the trigger is
set to Free Run
Configuring LXI Output of LAN Events via SCPI
Using SCPI to configure LXI Output LAN Events is more powerful than
using the front panel because there are more commands at the system
integrator's disposal. The following procedure represents the minimum
88
Action
Press Mode, Spectrum Analyzer
or IQ Analyzer (Basic) keys
Press the System hard key.
Press the softkeys I/O Config,
LXI, LXI Output LAN Events
Click on the event in the list that
should appear on the wire.
If LAN0-LAN7 were chosen in
Task 4, click the softkey labeled
"Source" then choose the desired
instrument event from the
menu.
If one of the last 5 events was
chosen, the identically named
instrument event has already
been selected as its source.
Click the one-of-n softkey
labeled "Enabled."
Press the hard key labeled
"Cont."
Press the Trigger hard key
followed by the Free Run
softkey.
Notes
You will see a list of event
names.
Notice that the last 5
events are named after
the instrument events
above.
This is the instrument
event that will cause a
peer to peer message to
go out on the line.
Chapter 3