Actual Load Calibration Procedure - Unipulse F370 Operation Manual

Digital indicator
Table of Contents

Advertisement

4.CALIBRATION

4-2. Actual Load Calibration Procedure

Follow the steps below to perform equivalent input calibration.
16
4-9.
Calibration
Protection Release
4-10.
Calibration
Value Selection
4-7.
Increment
4-8.
Decimal Place
4-3.
Zero Calibration
4-4.
Actual
Load Calibration
4-6.
Digital Offset
4-9.
Calibration Protection
Release the calibration protection.
Set the calibration value No.
(This step may be omitted for use with only one
selection.)
Enter the minimum value of digital increments.
(This step may be omitted if there is no change.)
Determine the decimal point place of the indicated value.
(This step may be omitted if there is no change.)
Set the zero point of the strain-gage sensor in no-load
condition (with the sensor unloaded).
Enter the span (gain) point of the strain-gage sensor with a
load applied to the sensor.
The calibrated value can be offset in advance.
(This step may be omitted if there is no change.)
Turn on the calibration protection for preventing
misoperation.