5-7. Contrast Adjustment

Adjust the contrast of the touch panel display.
Brightness can be adjusted by CONTRAST1.
Screen flickering can be adjusted by CONTRAST2.
How to set
Setting call

5-8. Backlight

The backlight is turned off if no button operation is performed for the set time (minutes).
The backlight is turned on by touching the panel. Setting 00 disables this function.
How to set
Setting call

5-9. Excitation Voltage

Select the bridge voltage to be supplied to the strain-gage sensor.
The bridge voltage is selectable from 10V, 5V and 2.5V.
Be sure to perform calibration after changing this setting.
How to set
Setting call
Operation Setting
Setting range: 00 - 99 min.
Operation Setting
Operation Setting
5.SETTING OF FUNCTIONS