7.HOLD FUNCTIONS

7-1-3. Auto Start Level

If you select the time specified period mode with trigger, or relative maximum, relative
minimum or inflection point hold in the hold mode setting, set the start level.
How to set

7-1-4. Hold Point Shift

In the "sample hold" and "inflection point hold", the sampling data is held retroactively
by the numerical value set under hold point shift.
How to set
46
Setting range: -99999 - 99999
Setting call
In each hold mode of relative maximum/relative minimum/inflection
point, if the auto start level is set at 99999, the hold operation starts
only with the period signal H/M without performing level detection.
Setting range: -99999 - 99999
w
When the hold point shift is set at 8,
the data eight pieces before
the actual hold point is held.
Caution
・This does not work for other holds.
・The retroactive sampling data is held irrespective of the hold
period. If the set numerical value is too large, point(s) out of the
period may be held.
Setting call
Page 1
Hold
Page 2
Hold
The actual hold point
t