6. COMPARISON FUNCTIONS
By the comparison function, the HI limit and LO limit values are set, and when the
indicated value exceeds the HI limit, the HI output is turned on, and when the indicated
value falls below the LO limit, the LO output is turned on. Also, HI-HI limit and LO-LO
limit values may be set outside the HI-LO limit comparison. When the indicated value
exceeds the HI-HI limit, the HH output is turned on, and when the indicated value falls
below the LO-LO limit, the LL output is turned on. When the HI, HH, LO and LL
outputs are all off, the OK output is turned on.
<HI/LO output conditions>
<HH/LL output conditions>
<OK output conditions>
HI-HI Limit
HI: indicated value > HI limit value
LO: indicated value < LO limit value
HH: indicated value > HI-HI limit value
LL: indicated value < LO-LO limit value
OK: All conditions of HH, HI, LO and LL are off.
Indicated value
Value
HI Limit
Value
LO Limit
Value
LL Limit
Value
OFF
HI
ON
OFF
LO
ON
OFF
HH
ON
OFF
LL
ON
OFF
OK
ON
6.COMPARISON FUNCTIONS
Time
37