4.CALIBRATION
4-6.

Digital Offset

By using the digital offset function, the value obtained by subtracting the set value from
the indicated value is displayed.
This function is convenient when zero cannot be obtained with no load for some reason
or for offsetting.
(displayed value)=(actual indicated value)-(offset value)
:
Setting range
- 99999 - 99999
How to set
Calibration
Setting call
Page 2
1)Press the MODE button.
2)Press the CALIBRATION button.
3)Press the PAGE button to select the page and press the Digital Offset button.
4) Enter the digital offset value by the numerical keys and determine with the OK button.