9-2-3. Level

Plotting is started when the conditions of the detection mode are met in comparison of
the trigger level set value and indicated value.
Plotting ends on one screen at the predetermined time of the X end point.
Trigger Level

9-2-4. External + Level

Plotting is started according to the conditions of the level detection mode in comparison
of the trigger level and indicated value after the ST/SP external input is turned on.
Plotting ends on one screen at the predetermined time of the X end point.
(Example) Level Detection Mode → Passing under.
Trigger Level
ST/SP
(Outside Input)
X end point time setteing
1.0msec more
X end point time setteing
Level Detection Period
9.WAVEFORM DISPLAY
Screen Hold、 Trigger wait
Plotting
Period
Plotting
Period
Screen Hold、 Trigger wait
t
t
71